A pupil dilator that is fast and easy to insert and remove, requires minimal intraocular manipulations, provides adequate pupil size and does not interfere with surgery.
Clinical Applications of the APX 200 Pupil Expanding Device
How to Use the APX 200 Pupil Expanding Device
APX 200 in Surgery
EXPANDER PLACEMENT
2
Perform another incision on the opposite side.
Fill the anterior chamber with a cohesive ophthalmic viscosurgical device (OVD). Inject additional OVDs between the iris and the anterior capsule to create space for the APX 200 terminal hooks.
4
Position the tips of the forceps gently as far as possible into the slots between the white base and the transparent cover to ensure that during engagement the tips of the forceps are properly positioned in relation to the Expander.
5
Using the forceps, make sure the Expander is fully closed to enable removal from the base.
6
The APX forceps are equipped with a sliding lock mechanism to enable advance locking of the preloaded forceps by the physician or assistant. The locking mechanism is operated when the forceps are fully closed by sliding the slider lock down to lock, up to unlock.
8
Partially open the Expander. Place the terminal hooks through the pupil under the iris and then open the Expander slowly. Do not completely release the Expander until the device is properly positioned.
9
Perform the same procedure on the opposite side.
10
11
Fully dilated pupil